(Sensors & Transducers (Canada) Via Acquire Media NewsEdge) Abstract: In this paper, an insole plantar pressure device based on 3-D forces piezoelectric sensor, integrated into a shoe, was developed for monitoring plantar pressure under real-life conditions. The device consisted of an insole with eight measure points composed of piezoelectric sensor, a wireless data transmission and embedded computer. The piezoelectric sensor with ceramic was embedded into the insole assembled by three different directions (3D) of X, Y and Z. The piezoelectric sensors mathematical model was built, and the layout of these sensors in insole was investigated. Based on the FFT algorithm, the eight measurement points data were collected and analyzed taken by frequency. The plantar changes of the left and right foot in 3D direction were measured, respectively. The results were displayed on the software interface, it showed that the insole system can be used to monitor plantar pressure during daily living and is expected to be useful in various clinical applications. Copyright © 2012 IPSA.



Keywords: Insole, Plantar pressure, Force piezoelectric sensor, Pressure measurement system.

(ProQuest: ... denotes formulae omitted.) 1. Instruction Insole pressure measurement systems have been widely used in clinical and research environments. Many studies have focused on plantar pressure measurements in healthy and hemiparetic adults or in healthy children, which can provide data that will optimize patient assessments and evaluate the treatment outcomes [1, 2]. To date, a number of devices have been developed to measure plantar pressure. However, there are significant variations in the data acquisition systems, such as the type of pressure sensors and the number and arrangement of the sensors used in the pressure measurement systems. Pressure measurement systems are commonly found in two different formats: an insole based or a platform based assessment system. An insole based pressure measurement systems compared with a platform based assessment system have some advantages, such as simple, convenience, real-time and so on. Therefore, they become good way and alternatives to evaluate plantar pressure real-time.


The F-Scan mobile system (Tekscan, Inc., Boston, USA) is one of the most commonly used in-sole pressure measurement system for gait analysis. Several studies have evaluated the accuracy and reliability of force and pressure measurements using the F-Scan®system and reached different conclusions. While the F-Scan system is less superior in comparison to another commercial product-the Pedar system, its accuracy and precision can be greatly improved by using appropriate pressure for calibration [12]. Pedar mobile system is a relatively new product among the insole pressure measuring devices. Although these devices have contributed to the basic analysis of human gait, some limitations have been noted from a therapeutic viewpiont. In particular, they are not designed for daily, real-time, and feedback use in therapy-directed research.

For piezoelectric sensors ceramic materials are used. The piezoelectric effect is found in non-conducting materials (e.g. quartz, ceramics, lead zirconate titanate) and in thin flexible PVDF films (polyviny lidenefluoride). The electronic dipoles in the material react under the influence of an external load with a displacement of charges on a molecular level generating electrical charges at the sensor surface. Charge amplifiers can be used to convert these charges to voltages. Kärki et al., have developed a new piezoelectric sensor prototype for plantar pressure measurements during gait. The mechanical stress at the plantar surface has two components, pressure acting normal to the surface and shear stress acting tangential to the surface [26]. Gross et al., have designed a system suitable for non-invasive measurement of discrete in-shoe vertical plantar stress during dynamic activities. In the system, eight transducers were constructed, with small piezoelectric ceramic squares (4.83 x 4.83 x 1,3 mm) used to generate a charge output proportional to vertical plantar stress [27].

Piezoelectric sensors (with the exception of PVDF) are highly elastic, show little material deformation and exhibit low hysteresis effects. Therefore, they are suitable for recording of high-frequency loading events. On the other hand PVDF and most piezoceramic materials are temperature sensitive so that the environmental conditions should be controlled and kept as constant as possible.

In this study, we developed an insole plantar measurement system. The piezoelectric sensors were distributed according to human anatomy characteristic. There were 8 measurement points in the insole. Each measurement point had three sensors assembled in 3D directions. Compared with previously literature reported plantar pressure measurement device, this insole device is smaller and lighter (the measurement unit was 10 mmx20 mmx8 mm in size, with a mass of 10 g, excluding the power source). Therefore, the pressure sensors can be fixed into the shoe insole, the device can be inserted into the subject's own shoes. Thus, the user wears one's own shoe, which is comfortable and does not interfere with the natural gait. The piezoelectric sensors mathematical model was built, and the layout of these sensors in insole was investigated. Based on the FFT algorithm, the eight measurement points data were collected and analyzed taken by frequency. The plantar changes of the left and right foot in 3D direction were measured, respectively. The results showed that this insole plantar pressure measurement system can be used to monitor plantar pressure during daily living.

2. Device Design 2.1. Piezoelectric Sensors Mathematical Model As shown in Fig. 1, pressure sensor can be viewed as a mechanical system, and it is a quality-springdamper system, belonging to the second sensor.

Its differential equation is: ... (1) It can be rewritten as: ... (2) where m is the portion of its mass during system movement, c is the damping coefficient of this system, k is the stiffness of the spring, con is the natural frequency of this system, and here, con = dk/m * C is the damping ratio of this system.

Equation 1 can be expressed with the second of differential equation, which is following: ... (3) It can be rewritten as : ... (4) when k=b0/a0 is the static sensitivity; con = yfaja^ is the undamped natural frequency: ç = aj2^ja0a2 is the damping ratio, D is differential operator, ai, a2, ao, bo are the system constant, usually, we take ao^3o=l.

Therefore, the transfer function of the second system is following: ... (5) Let ...

Therefore, the transfer function changed into: ... (6) where ... and ... are viewed as two major parameters in the piezoelectric sensor.

The frequency response function of frequency characteristics of the second order system is following: ... (7) Amplitude frequency characteristic is the following: ... (8) Phase frequency characteristic is the following: ... (9) As above mentioned, after the known transfer function, we can easily find out the characteristics of the system parameters con and Ç , the frequency response function, the amplitude frequency and phase frequency characteristics can be gained. Therefore, we can analyze the difference between output and input to reduce the dynamic error. Therefore, we can select the suitable vale of Ç and ... to ascertain the scope of frequency.

Usually, ... is in the flat section of the amplitude frequency curve.

2.2. Pressure Sensor's Layout in an Insole Feet, as the body's base of support, often endure ground reaction forces during daily activities. As shown in Fig. 2, anatomical structure of foot: HL, heel lateral (1,3); HM, heel medial (2); M5, metatarsal 5 (4); M4 metatarsal 4 (5); M3, metatarsal 3 (6); M2, metatarsal 2 (7); Ml, metatarsal 1 (8). The purpose of assessing peak plantar pressures during dynamic gait was to quantify the highest pressures applied to each region of the plantar surface of the subject's feet as they occur at any point during foot contact in typical activities of daily living. The entire sensors are inserted into two insoles equally. Each insole has 8 measurement points, its distribution was shown in Fig. 2.

2.3. Eight Measurement Points Data Collection Based on the FFT algorithm, the eight measurement points data were collected and analyzed taken by frequency. The FFT algorithm was shown in Fig. 3.

The 8 measurement points data wre showm on the software interface (Fig. 4). The user can observe each measurement point onsite by the software interface.

3. Materials and Methods 3.1. Subjects Three healthy young subjects (male, 20 years old, body weight: 65 kg) participated in this experiment. All subjects participated in plantar measurement after providing informed consent for experimental procedures.

3.2. Experimental Protocol Each subject wore the device and performed stand and straight walking trial at a comfortable speed for 10 m. The plantar pressure during stand and straight walking were measured continuously. The pressure data during transitional gait (i.e., the first 4 steps and the last 4 steps) were excluded from analysis, so only data obtained during steady gait were analyzed during straight walking.

4. Results and Discussion Based on the FFT algorithm, the eight measurement points data were collected and analyzed taken by frequency. Table 1 and Table 2 were the plantar measurement data about right foot and left foot during the subjects were standing posture, the results showed that the pressure of Z direction were the maximum and the pressure of X direction were the minimum at left and right foot plantar pressure same measurement point, The forefoot fifth and third metatarsal head region of pressure were the maximum pressure. The measurement data were conducted in accordance with the previous literature reported conclusions [22-23]. The results also showed that the proposed plantar pressure measurement system based on the 3D piezoelectric sensor can satisfy the actual plantar pressure measurement during daily human activity.

5. Conclusion In the present study, we developed an insole plantar pressure device based on 3-D forces piezoelectric sensor, integrated into a shoe for monitoring plantar pressure under real-life conditions. The device consisted of an insole with eight measure points composed of piezoelectric sensor, a wireless data transmission and embedded computer. The piezoelectric sensors were distributed according to human anatomy characteristic. There were 8 measurement points in the insole. Each measurement point had three sensors assembled in 3D directions. Because pressure sensors are fixed into the shoe insole, the device can be inserted into the subject's own shoes. Thus, the user wears one's own shoe, which is comfortable and does not interfere with the natural gait. The insole device is small and light (the measurement unit was 10 mmx20 mmx8 mm in size, with a mass of 10 g, excluding the power source). The piezoelectric sensors mathematical model was built. Based on the FFT algorithm, the eight measurement points data were collected and analyzed taken by frequency. The plantar changes of the left and right foot in 3D direction were measured, respectively. The results showed that the insole system can be used to monitor plantar pressure during daily living and is expected to be useful in various clinical applications.

References [1]. R. Naemi, T. L. Chevalier, A. Healy, N. Chockalingam, The effect of the use of a walkway and the choice of the foot on plantar pressure assessment when using pressure platforms, The Foot, Vol. 22, Issue 2,2012, pp. 100-104.

[2]. P. A. Deleu, T. Leemrijse, B. Vandeleene, P. Maldague, B. Devos Bevemage, Plantar pressure relief using a forefoot offloading shoe, Foot and Ankle Surgery, Vol. 16, Issue 4, 2010, pp. 178-182.

[3]. K. E. Fiedler, W. J. A. Stuijfzand, J. Harlaar, J. Dekker, H. Beckerman, The effect of shoe lacing on plantar pressure distribution and in-shoe displacement of the foot in healthy participants, Gait & Posture, Vol. 33, Issue 3,2011, pp. 396-400.

[4]. P. R. Cavanagh, F.G. Hewitt Jr, J.E. Perry, In-shoe plantar pressure measurement: a review, The Foot, Vol. 2, Issue 4, 1992, pp. 185-194.

[5]. Z. Sawacha, G. Guameri, G. Cristoferi, A. Guiotto, A. Avogaro, C. Cobelli, Integrated kinematicskinetics-plantar pressure data analysis: A useful tool for characterizing diabetic foot biomechanics, Gait & Posture, Vol. 36, Issue 1,2012, pp. 20-26.

[6]. Z. Pataky, L. Faravel, J. D. Silva, J. Assal, A new ambulatory foot pressure device for patients with sensory impairment. A system for continuous measurement of plantar pressure and a feed-back alarm, Journal of Biomechanics, Vol. 33, Issue 9, 2000, pp. 1135-1138.

[7]. T. L. Chevalier, H. Hodgins, N. Chockalingam, Plantar pressure measurements using an in-shoe system and a pressure platform: A comparison, Gait & Posture, Vol. 31, Issue 3,2010, pp. 397-399.

[8]. K. S. Maluf, R. E. Morley Jr, E. J. Richter, J. W. Klaesner, M. J. Mueller, Monitoring in-shoe plantar pressures, temperature, and humidity: Reliability and validity of measures from a portable device, Archives of Physical Medicine and Rehabilitation, Vol. 82, Issue 8,2001, pp. 1119-1127.

[9]. M. J. Mueller, M. J. Strube, Generalizability of in-shoe peak pressure measures using the F-scan system, Clinical Biomechanics, Vol. 11, Issue 3, 1996, pp. 159-164.

[10]. P. Catalfamo, D. Moser, S. Ghoussayni, D. Ewins, Detection of gait events using an F-Scan in-shoe pressure measurement system, Gait & Posture, Vol. 28, Issue 3,2008, pp. 420-426.

[11]. M. Brown, S. Rudicel, A. Esquenazi, Measurement of dynamic pressure at the shoe-foot interface during normal walking with various foot orthoses using the F-Scan system, Foot and Ankle International, Vol. 17, Issue 5,1996, pp. 152-156.

[12]. H. L. P. Hurkmans, J. B. J. Bussmann, E. Benda, J. A. N. Verhaar, H. J. Siam, Accuracy and repeatability of the Pedar Mobile system in long-term vertical force measurements, Gait & Posture, Vol. 23, Issue 1,2006, pp. 118-125.

[13]. A. K. Ramanathan, P. Kiran, G. P. Arnold, W. Wang, R. J. Abboud, Repeatability of the Pedar-X® in-shoe pressure measuring system, Foot and Ankle Surgery, Vol. 16, Issue 2,2010, pp. 70-73.

[14]. J. E. Perry, J. O. Hall, B. L. Davis, Simultaneous measurement of plantar pressure and shear forces in diabetic individuals, Gait & Posture, Vol. 15, Issue 1, 2002, pp. 101-107.

[15]. M. Saito, K. Nakajima, C. Takano, Y. Ohta, C. Sugimoto, R. Ezoe, K. Sasaki, H. Hosaka, T. Ifukube, S. Ino, K. Yamashita, An in-shoe device to measure plantar pressure during daily human activity, Medical Engineering & Physics, Vol. 33, Issue 5, 2011, pp. 638-645.

[16]. H. L. Hurkmans, J.B. Bussmann, E. Benda, J. A. Verhaar, H. J. Siam, Techniques for measuring weight bearing during standing and walking, Clinical Biomechanic, Vol. 18, Issue 7,2003, pp. 576-589.

[17]. H. J. Hermie, C. A. Waal, J. Buurke, G. Zilvold, A new gait analysis system for clinical use in a rehabilitation center, Orthopedics, Vol. 9, Issue 12, 1986, pp. 1669-1675.

[18]. P. Moretto, M. Bisiaux, M. A. Lafortune, Froude number fractions to increase walking pattem dynamic similarities: Application to plantar pressure study in healthy subjects, Gait & Posture, Vol. 25, Issue 1, 2007, pp. 40-48.

[19]. H. B. Menz, M. E. Morris, Clinical determinants of plantar forces and pressures during walking in older people, Gait & Posture, Vol. 24, Issue 2, 2006, pp .229-23 6.

[20]. J. M. Bumfield, C. D. Few, O. S. Mohamed, J. Perry, The influence of walking speed and footwear on plantar pressures in older adults, Clinical Biomechanics, Vol. 19, Issue 1,2004, pp. 78-84.

[21]. J. S. Akins, K. A. Keenan, T. C. Sell, J. P. Abt, S. M. Lephart, Test-retest reliability and descriptive statistics of geometric measurements based on plantar pressure measurements in a healthy population during gait, Gait & Posture, Vol. 35, Issue 1, 2012, pp. 167-169.

[22]. D. Rosenbaum, H. P. Becker, Plantar pressure distribution measurements. Technical background and clinical applications, Foot and Ankle Surgery, Vol. 3, Issue 1,1997, pp.1-14.

[23]. J. Perin, G. Agnesina, R. Taiar, Y. Toshev, Analysis of the plantar foot pressure during walking: VFE position and VRI position, Journal of Biomechanics, Vol 39, Supplement 1,2006, pp. S552-S553.

[24]. K. F. Lei, K. F. Lee, M. Y. Lee, Development of a flexible PDMS capacitive pressure sensor for plantar pressure measurement, Microelectronic Engineering, In Press, 2012.

[25]. K. M. Leitch, T. B. Birmingham, Ian C. Jones, J. Robert Giffin, Thomas R. Jenkyn, In-shoe plantar pressure measurements for patients with knee osteoarthritis: Reliability and effects of lateral heel wedges, Gait & Posture, Vol. 34, Issue 3, 2011, pp. 391-396.

[26]. S. Kärki, J. Lekkala, H. Kuokkanen, J. Halttunen, Development of a piezoelectric polymer film sensor for plantar normal and shear stress measurements, Sensors and Actuators A: Physical, Vol. 154, Issue 1, 2009, pp. 57-64.

[27]. T. S. Gross, R. P. Bunch, Measurement of discrete vertical in-shoe stress with piezoelectric transducers, Journal of Biomedical Engineering, Vol. 10, Issue 3, 1988, pp. 261-265 [28]. J. Hughes, P. Clark, R. R. Jagoe, C. Gerber, L. Klenerman, The pattem of pressure distribution under the weightbearing forefoot, The Foot, Vol. 1, Issue 3, 1991, pp. 117-124.

[29]. J. K. Gurney, U. G. Kersting, D. Rosenbaum, Between-day reliability of repeated plantar pressure distribution measurements in a normal population, Gait & Posture, Vol. 27, Issue 4,2008, pp. 706-709.

[30]. K. M. Shah, M. J. Mueller, Effect of selected exercises on in-shoe plantar pressures in people with diabetes and peripheral neuropathy, The Foot, in Press, Corrected Proof, 2012.

[31]. K J. Chesnin, L. Selby-Silverstein, M. P. Besser, Comparison of an in-shoe pressure measurement device to a force plate: concurrent validity of center of pressure measurements, Gait & Posture, Vol. 12, Issue 2,2000, pp. 128-133.

[32]. N. A. Guidemond, P. Leffers, N.C. Schaper, A. P. Sanders, F. Nieman, P. Willems, G. H. I. M. Walenkamp, The effects of insole configurations on forefoot plantar pressure and walking convenience in diabetic patients with neuropathic feet, Clinical Biomechanics, Vol. 22, Issue 1,2007, pp. 81-87.

[33]. J. W. K. Tong, E. Y. K. Ng, Preliminary investigation on the reduction of plantar loading pressure with different insole materials (SRP - Slow Recovery Poron®, P - Poron®, PPF - Poron® + Plastazote, firm and PPS - Poron® + Plastazote, soft) Spatial resolution in plantar pressure measurement, Journal of Biomechanics, 22 June, 2012.

[34]. J. J. Wersch, J. G. Webster, W. J. Tompkins, A portable insole plantar pressure measurement system. J. Rehabil. Res. Dev., Vol. 29, Issue 1, 1992, pp. 13-18.

[35]. A. Authier, P. Boissy, J. Beauvais, R. Fontaine, A proof of concept for a wireless ambulatory weight bearing measurement system in rehabilitation and telerehabilitation applications, in Proceedings of 3rd IEEE International Conference on Wireless and Mobile Computing, Networking and Communications, 2007, pp. 73.

[36]. A. M. Nova, J. C. Cuevas-García, R. Sánchez-Rodríguez, J. Pascual-Huerta, E. Sánchez-Barrado, Study of plantar pressure patterns by means of instrumented insoles in subjects with hallux valgus, Revista Española de Cirugía Ortopédica y Traumatología (English Edition), Vol. 52, Issue 2,2008, pp. 94-98.

[37]. D. Santos, T. Carline, R. Richmond, R. J. Abboud, A modular device to measure the effects of plantar foot pressure on the microcirculation of the heel, The Foot, Vol 13, Issue 1,2003, pp. 30-38.

[38]. M. Lord, Spatial resolution in plantar pressure measurement, Medical Engineering & Physics, Vol. 19, Issue 2, 1997, pp. 140-144.

[39]. H. Jonely, J. M. Brismée, P. S. Sizer Jr., C. R. James, Relationships between clinical measures of static foot posture and plantar pressure during static standing and walking, Clinical Biomechanics, Vol. 26, Issue 8, 2011, pp. 873-879.

[40]. N. L. W. Keijsers, N. M. Stolwijk, B. Nienhuis, J. Duysens, A new method to normalize plantar pressure measurements for foot size and foot progression angle, Journal of Biomechanics, Vol. 42, Issue 1,2009, pp. 87-90.

[41]. P. W. Kong, H. D. Heer, Wearing the F-Scan mobile in-shoe pressure measurement system alters gait characteristics during running, Gait & Posture, Vol. 29, Issue 1,2009, pp. 143-145.

[42]. L. Begg, J. Bums, A comparison of insole materials on plantar pressure and comfort in the neuroischaemic diabetic foot, Clinical Biomechanics, Vol. 23, Issue 5, 2008, pp. 710-711.

[43]. V. Tessutti, F. T. Souza, A. P. Ribeiro, A. L. Nunes, I. de Camargo Neves Sacco, In-shoe plantar pressure distribution during running on natural grass and asphalt in recreational runners, Journal of Science and Medicine in Sport, Vol. 13, Issue 1, 2010, pp. 151-155.

Jicheng ZHAO, Yemin GUO, Lanmei WANG School of Agriculture and Food Engineering, Shandong University of Technology, No. 12, Zhangzhou Road, Zibo 255049, Shandong Province, P. R. China Tel.:+86-533-2786702, fax: +86-533-2786702 E-mail: [email protected] Received: 17 October 2013 /Accepted: 22 November 2013 /Published: 30 December 2013 (c) 2013 International Frequency Sensor Association