Defibtech Announces a Voluntary Recall of DDU-100 series AEDs
GUILFORD, Conn. --(Business Wire)--
Defibtech, LLC is initiating a worldwide voluntary recall of certain
DDU-100 series semi-automatic external defibrillators (AEDs) sold under
the Lifeline AED and ReviveR AED brand names, including 65,885 AEDs
distributed in the United States. This recall affects only DDU-100
Series AEDs shipped with 2.004 software or earlier. This corrective
action addresses two possible conditions, which in rare cases may cause
an affected AED to cancel shock during the charging process and not
provide therapy which may result in failure to resuscitate the patient.
Both conditions are not detectable by the periodic self test.
Condition 1: In rare instances, the AED may cancel charge in preparation
for a shock. Based on field data, the odds of an affected AED having
this happen are less than a 1 in 400,000 chance per month for any given
AED.
A subset of AEDs (less than 11%) that are affected by Condition 1 are
also affected by Condition 2: In rare instances, the AED may cancel
charge in preparation for a shock in very high humidity conditins. The
only reported cases were in environments of greater than 95% relative
humidity or condensing conditions. Based on field data, the odds of an
affected AED having this happen are less than a 1 in 250,000 chance per
month for any given affected AED.
Defibtech will provide customers with a free software upgrade to address
these issues. The correction to the AED will be able to be performed at
the location where the AED is deployed.
Because both of these conditions occur very rarely, it is recommended
that customers keep their AEDs in service until they have performed the
software upgrade. Full instructions and recommendations are being mailed
to affected customers. This customer notification, as well as
instructions on determining whether an AED is affected, can also be
found on the www.defibtech.com/fa11
web page. For additional information regarding this recall, please refer
to the above referenced web page, contact your distributor, or contact
Defibtech at techsupport@defibtech.com,
1-877-453-4507 or 1-203-453-4507.
The AEDs affected by this recall have been distributed globally to fire
departments, EMS, health clubs, schools, and other organizations. The
Food and Drug Administration (FDA) has determined that this action is a
Class I recall. Any adverse reactions experienced with the use of this
product and/or quality problems should also be reported to the FDA's
MedWatch Program by phone at 1-800- FDA-1088, or on the MedWatch website
at www.fda.gov/medwatch.